Agilent Technologies 87020 Stroller User Manual


 
4-43
Performance Tests
11. System Trace Noise (Only for Analyzers with Option 006)
3
Press
MARKER FCTN
,
MARKER MODE MENU
,
STATS ON
to activate the instruments
statistic feature.
System Trace Noise for A/R Magnitude from 30 kHz to 3 GHz
4
Press
MEAS
,
Trans: E/E S12 REV
.
5
Press
MENU
,
CW FREQ
,
3
,
G/n
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
6
When the analyzer finishes the number of sweeps, press
SCALE REF
,
AUTO SCALE
.
7
Write the s.dev (standard deviation) value shown, which appears on the
analyzer display, on the Performance Test Record.
System Trace Noise for A/R Magnitude from 3 GHz to 6 GHz
8
Press
MENU
,
CW FREQ
,
6
,
G/n
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
9
When the analyzer finishes the number of sweeps, press
SCALE REF
,
AUTO SCALE
.
10
Write the s.dev (standard deviation) value, which appears on the analyzer
display, on the Performance Test Record.
System Trace Noise for A/R Phase from 3 GHz to 6 GHz
11
Press
FORMAT
,
PHASE
.
12
Press
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
13
When the analyzer finishes the number of sweeps, press
SCALE REF
,
AUTO SCALE
.
14
Write the s.dev (standard deviation) value, which appears on the analyzer
display, on the Performance Test Record.
System Trace Noise for A/R Phase from 30 kHz to 3 GHz
15
Press
MENU
,
CW FREQ
,
3
,
G/n
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
16
When the analyzer finishes the number of sweeps, press
SCALE REF
,
AUTO SCALE
.
17
Write the s.dev (standard deviation) value, which appears on the analyzer
display, on the Performance Test Record.
System Trace Noise for B/R Magnitude from 30 kHz to 3 GHz
18
Press
MEAS
,
Trans: E/E S21 FWD
,
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1.
19
When the analyzer finishes the number of sweeps, press
SCALE REF
,
AUTO SCALE
.
20
Write the s.dev (standard deviation) value, which appears on the analyzer