Agilent Technologies 87020 Stroller User Manual


 
4-30
Performance Tests
8. Test Port Crosstalk
Required
Equipment
For 50 Analyzers
Calibration Kit, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85031B
Cable, APC-7 24-inch . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent P/N 8120-4779
For 75 Analyzers
Calibration Kit, 75
, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85036B
Procedure
1
Connect the equipment as shown in Figure 4-18.
Figure 4-18. Test Port Crosstalk Test Setup
2
Press
PRESET
,
MENU
,
POWER
,
10
,
x1
.
3
Press
AVG
,
IF BW
,
10
,
x1
.