Automated Verification Option 011
Step 3. Device Verification
18
Press
TRANSMISSION, FWD TRANS THRU
.
19
When the analyzer finishes the measurement, press
FWD MATCH THRU
.
20
When the analyzer finishes the measurement, press
REV TRANS THRU
.
21
When the analyzer finishes the measurement, press
REV MATCH THRU
.
22
When the analyzer displays,
PRESS DONE IF FINISHED WITH STD(s)
,
press
STANDARDS DONE, DONE 2-PORT CAL
.
23
Press
SAVE/RECALL
,
SELECT DISK, INTERNAL MEMORY, RETURN, SAVE STATE
to save
the calibration into the analyzers internal memory.
24
When the analyzer finishes storing the measurement calibration, press
SELECT
DISK
,
INTERNAL DISK
.
Step 3. Device Verification
1
Press
SYSTEM
,
SERVICE MENU, TESTS,
28
,
x1
,
EXECUTE TEST
.
2
At the prompt, connect the 20 dB attenuator (supplied in the verification kit)
as shown in Figure 5-4.
Figure 5-4. Connections for the 20 dB Verification Device