Performance Tests Option 011
6. Receiver Channel Noise Floor Level (Only for Analyzers with Option 006)
Receiver Channel A Noise Floor Level with 10 Hz IF BW
(3 GHz6 GHz)
33
Press
MEAS
,
INPUT PORTS
,
A
.
34
Press
MENU
,
TRIGGER MENU
,
SINGLE
.
35
When the analyzer finishes the sweep, record the mean value, which appears
on the analyzer display.
36
Use the following equation to convert the linear magnitude mean value to log
magnitude:
Power (dBm) = 20 * [log
10
(linear magnitude mean value)]
37
Write this calculated value in the Performance Test Record in dBm.
Receiver Channel A Noise Floor Level with 3 kHz IF BW
(3 GHz6 GHz)
38
Press
AVG
,
IF BW
,
3
,
k/m
.
39
Press
MENU
,
TRIGGER MENU
,
SINGLE
.
40
When the analyzer finishes the sweep, record the mean value, which appears
on the analyzer display.
41
Use the following equation to convert the linear magnitude mean value to log
magnitude:
Power (dBm) = 20 * [log
10
(linear magnitude mean value)]
42
Write this calculated value in the Performance Test Record in dBm.
In Case of
Difficulty
1
Perform the ADC Offset Correction Constants (Test 52), located in the
Adjustments chapter in the
Agilent 8753D Option 011 Service Guide
.
2
Suspect the A10 digital IF assembly if
both
receiver channels fail.
3
Refer to the
Agilent 8753D Option 011 Service Guide
for further
troubleshooting information.