6-63
Performance Tests Option 011
16. Source and Receiver Harmonics (Option 002 Only)
Figure 6-32. Source Harmonics Test Setup
2
Press
PRESET
,
MENU
,
POWER
,
20
,
x1
.
3
Press
START
,
16
,
M/
µ
,
STOP
,
1.5
,
G/n
to set the frequency range for searching the
source 2nd harmonic.
4 Only for Analyzers with Option 006:
Press
STOP
,
3
,
G/n
.
5
Press
AVG
,
IF BW
,
10
,
x1
to set the IF bandwidth to 10 Hz.
6
Press
MEAS
,
INPUT PORTS
,
A
.
7
After one sweep, press
DISPLAY
,
DATAMEMORY
,
DATA/MEM
to normalize the
trace and record the fundamental.
8
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC SECOND
to prepare the analyzer for
making a 2nd harmonic measurement.
9
After one sweep, press
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH
MAX
.
10
Write the marker 1 value (which appears on the analyzer display) on the
"Performance Test Record." This is the worst case source 2nd harmonic.
11
Press
STOP
,
1
,
G/n
to change the stop frequency to 1 GHz.
12 Only for Analyzers with Option 006:
Press
STOP
,
2
,
G/n
.