5-17
Automated Verification Option 011
Step 3. Device Verification
as shown in Figure 5-11.
Figure 5-11. Connections for the 20 dB Attenuator
3
Press
CONTINUE
to run the test:
If you switched the record function
off
, you have to press
CONTINUE
after
each S-parameter measurement.
If you switched the record function
on
, the test displays and prints the pass/
fail information for the S-parameter measurements that are valid for system
verification.