4-20
Performance Tests
7. Test Port Input Frequency Response
Figure 4-10. Setup for Power Meter Calibration on Test Port 1
3
Press
PRESET
,
START
,
300
,
k/m
.
4 Only for Analyzers with Option 006:
Press
STOP
,
3
,
G/n
.
5
Press
LOCAL
,
SYSTEM CONTROLLER
.
6
Press
SET ADDRESSES
and
POWER MTR
until the analyzer shows the correct power
meter model.
7
Press
ADDRESS: P MTR/HPIB
. The default power meter GPIB address is 13. Make
sure it is the same as your power meter GPIB address. Otherwise, use the
analyzer front panel keypad to enter the correct GPIB address for your power
meter.