Performance Tests
3. Test Port Output Power Accuracy
5
Check to see if the analyzer is phase locking to the external CW signal:
If the analyzer displays any phase lock error messages, write unlock in the
Performance Test Record for the set CW signal.
If the analyzer does
not
display any phase lock error messages, write lock
in the Performance Test Record for the set CW signal.
6
On the external source, press
CW
,
50
,
MHz/
µ
.
7
On the analyzer, press
50
,
M/
µ
.
8
Repeat Step 5 through Step 7 for the other external source CW frequencies
listed in the Performance Test Record. For analyzers with Option 006, change
the power sensor to the Agilent 8481A for frequencies above 4 GHz.
In Case of
Difficulty
1
Be sure the external source power is set within 0 to 25 dBm.
2
Make sure the analyzers Ext Source Auto" feature is selected. In addition,
verify that the analyzer is set to measure its input channel R.
3
Verify that all connections are tight.
3. Test Port Output Power Accuracy
Perform this test to confirm the accuracy of the Agilent 8702D source output
power.
Required
Equipment
Equipment Required for 50 Analyzers
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8482A
Adapter, APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11524A
Additional Equipment Required for Analyzers with Option 006
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8481A
Equipment Required for 75 Analyzers
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8483A Option H03