6-30
Performance Tests – Option 011
9. Receiver Input Crosstalk
(Only for Option 006 Analyzers) B into A Crosstalk from 4.5 GHz to
6 GHz
30
Press
START
,
4.5
,
G/n
,
STOP
,
6
,
G/n
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
31
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”
A into B Crosstalk from 300 kHz to 1 GHz
32
Connect the equipment as shown in Figure 6-15.
Figure 6-15. A into B Receiver Input Crosstalk Test Setup
33
Press
START
,
300
,
k/m
,
STOP
,
1
,
G/n.
34
Press
MEAS
,
INPUT PORTS
,
A/R
,
AVG
,
AVERAGING OFF
,
MENU
,
TRIGGER MENU
,
SINGLE
.
35
At the end of the sweep, press
DISPLAY
,
DATA→MEMORY
,
DATA/MEM
,
MEAS
,
INPUT
PORTS
,
B/R
.
36
Press
AVG
,
AVERAGING FACTOR
,
5
,
x1
,
AVERAGING ON
,
MENU
,
TRIGGER MENU
,
NUMBER
of GROUPS
,
5
,
x1
.
37
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”