6-44
Performance Tests Option 011
12. Receiver Magnitude Dynamic Accuracy
a
Press
CAL
,
CALIBRATE MENU
,
RESPONSE
,
THRU
.
b
After the analyzer finishes the measurement, press
DONE: RESPONSE
.
c
Press
SAVE/RECALL
,
SAVE STATE
.
22
Set the Agilent 8496A to 0 dB.
23
On the Agilent 8702D Option 011, press
MENU
,
TRIGGER MENU
,
SINGLE
.
24
Write the mean value in the "Receiver Channel B" column of the "Performance
Test Record."
25
Write the Agilent 8902A readout in the "8902A" column of the "Performance
Test Record."
26
Calculate the difference between the "Receiver Channel B" and the "8902A".
27
Take the absolute value of the calculated difference and enter the result in the
"Magnitude Value" column of the "Performance Test Record."
28
Set the Agilent 8496A to the other attenuation settings listed in the
"Performance Test Record."
29
On the Agilent 8702D Option 011, press
SINGLE
. Repeat Step 23 through Step
27.
Channel R Magnitude Dynamic Accuracy
30
Connect the equipment as shown in Figure 6-24.