Agilent Technologies 87020 Stroller User Manual


 
3-6
Manual Verification
Step 3. Device Verification
Figure 3-3.
Figure 3-3. Connections for the 20 dB Verification Device
3
Press
CONTINUE
to run the test.
4
Press
MENU
,
POWER
,
–2
,
x1
,
CAL
,
CORRECTION ON
.
5
Press
COPY
,
PRINT MONOCHROME
to obtain the graphical representation of the
measurement.
6
Press
COPY
,
LIST
,
PRINT ALL MONOCHROME
if you prefer a tabular representation
of the measurement.
7
Press
SYSTEM
,
SERVICE MENU
,
TEST OPTIONS
,
CONTINUE TEST
.
Ignore the error message,
CAUTION: CALIBRATION REQUIRED
.
8
Press
MENU
,
POWER
,
–2
,
x1
,
CAL
,
CORRECTION ON
.
9
Repeat Step 5 through Step 8 until you have made all the S-parameter
measurements for the verification device.
For each verification device, there are a total of eight measurements (magni-
For those S-parameter measurements with upper and lower point limits shown on the
Agilent 8702D display, you will notice that when you press the
CORRECTION
ON
softkey,
the measurement test status changes from
FAIL
to
PASS
.
If the test status remains
FAIL
after you press
CORRECTION
ON
, refer to “In Case of
Difficulty” on page 3-9.