4-56
Performance Tests
14. Test Port Receiver Phase Dynamic Accuracy
Figure 4-33. Dynamic Accuracy (Phase)
Procedure Test Port 2 Phase Dynamic Accuracy
1
Refer to the Test Port Receiver Magnitude Dynamic Accuracy portion of the
Performance Test Record, for the values in the Calculated Value column.
2
Copy the values in the Magnitude Value column to the Calculated Value
column of the Test Port Receiver Phase Dynamic Accuracy portion of the
Performance Test Record.
3
For every test port input power level, calculate A using the following formula:
4
Write the result in the A column of the Performance Test Record.
5
For every test port input power level, calculate B using the following formula:
B = 1 A
6
Write the result in the B column of the Performance Test Record.
A 10
CalculatedValue
20
---------------------------------------------


=