Agilent Technologies 87020 Stroller User Manual


 
4-10
Performance Tests
5. Minimum R Channel Level
6
Write the power meter reading in the "Results Measured" column on the
"Performance Test Record."
7
Calculate the difference between the analyzer test port power (which appears
on the analyzers display) and the power meter reading. Write the result in the
"Power Level Linearity" column on the "Performance Test Record."
8
Repeat Step 5 through Step 7 for the other power levels listed in the
"Performance Test Record."
9
Press
CW FREQ
,
3
,
G/n
.
10
Set the power meter calibration factor for this CW frequency and press
REL
to
set the reference at this new frequency.
11
Press
MENU
,
POWER
,
15
,
x1
.
12
Write the power meter reading in the "Results Measured" column on the
"Performance Test Record."
13
Calculate the difference between the analyzer test port power and the power
meter reading. Write the result in the "Power Level Linearity" column of the
"Performance Test Record."
14
Repeat Step 11 through Step 13 for the other power levels listed in the
"Performance Test Record."
In Case of
Difficulty
1
Ensure that the power meter and power sensor(s) are operating to
specifications. Be sure you set the power meter calibration factor for the CW
frequency that you are testing.
2
Verify that there is power coming out of the analyzers test port 1. Be sure you
did
not
accidently switch off the analyzers internal source. If you did so, press
MENU
,
POWER
,
SOURCE PWR ON
.
3
Repeat this performance test.
5. Minimum R Channel Level
Perform this test to determine the minimum R channel input power level at
which phase lock can be accomplished.