Agilent Technologies 87020 Stroller User Manual


 
4-69
Performance Tests
18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only)
27
Press
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH: MAX
.
28
Write the marker 1 value (which appears on the analyzer display) on the
Performance Test Record. This is the worst case port 1 input (receiver
channel A) 2nd harmonic.
Port 1 Input Worst Case 3rd Harmonic
29
Press
STOP
,
2
,
G/n
to change the stop frequency for measuring the receiver 3rd
harmonic.
30
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
.
31
After one sweep, press
DISPLAY
,
DATAMEMORY
,
DATA/MEM
to normalize the
trace.
32
Press
SCALE REF
,
AUTO SCALE
,
SCALE/DIV
,
1
,
x1
to get a better viewing of the trace.
33
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC THIRD
.
34
After one sweep, press
SCALE REF
,
AUTO SCALE
.
35
Press
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH:MAX
.
36
Write the marker 1 value on the Performance Test Record.
37
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
.
Port 2 Input Worst Case 2nd Harmonic
38
Press
STOP
,
3
,
G/n
to set the stop frequency for measuring the 2nd harmonic.
39
Press
MEAS
,
Trans: E/E S21 FWD
,
INPUT PORTS
,
B
.
40
After one sweep, press
DISPLAY
,
DATAMEMORY
,
DATA/MEM
to normalize the
trace.
41
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC SECOND
.
42
After one sweep, press
SCALE REF
,
AUTO SCALE
to get a better viewing of the
trace.
43
Press
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH MAX
.
44
Write the marker 1 value (which appears on the analyzer display) on the
Performance Test Record. This is the worst case port 2 input (receiver
channel B) 2nd harmonic.
Port 2 Input Worst Case 3rd Harmonic
45
Press
STOP
,
2
,
G/n
to change the stop frequency for measuring the receiver 3rd