Performance Tests
4. Test Port Output Power Range and Linearity
service manual for more information on how to do this task.
2
On the analyzer, press
PRESET
,
MENU
,
CW FREQ
,
300
,
k/m
. Set the power meter
calibration factor for this CW frequency.
3
Connect the equipment as shown in Figure 4-4.
Figure 4-4. Test Port Output Power Range and Accuracy Test Setup
4
On the Agilent 438A, press
REL
. This sets the current power level for relative
power measurement.
5
On the analyzer, press
MENU
,
POWER
,
PWR RANGE MAN
.