5-16
Automated Verification Option 011
Step 3. Device Verification
10
When the analyzer finishes measuring the standard, press
STANDARDS DONE
.
11
Connect the test port cables as shown Figure 5-10.
Figure 5-10. Transmission Calibration Setup
12
Press
TRANSMISSION, FWD TRANS THRU
.
13
When the analyzer finishes the measurement, press
FWD MATCH THRU
.
14
When the analyzer displays,
PRESS DONE IF FINISHED WITH STD(s)
,
press
STANDARDS DONE, DONE 2-PORT CAL
.
15
Press
SAVE/RECALL
,
SELECT DISK, INTERNAL MEMORY, RETURN, SAVE STATE
to save
the calibration into the analyzers internal memory.
16
When the analyzer finishes storing the measurement calibration, press
SELECT
DISK
,
INTERNAL DISK
.
Step 3. Device Verification
1
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
28
,
x1
,
EXECUTE TEST
.
2
At the prompt, connect the 20 dB attenuator (supplied in the verification kit)