2-5
Automated Verification
Step 2. Measurement Calibration
Figure 2-2. Connections for Measurement Calibration Standards
5
Press
FORWARD:OPEN
.
6
When the Agilent 8702D finishes measuring the standard, connect the short
end of the open/short combination to reference test port 1.
7
Press
FORWARD:SHORT
.
8
When the Agilent 8702D finishes measuring the standard, connect the 50 ohm
termination (supplied in the calibration kit) to reference test port 1.
9
Press
FORWARD:LOAD
.
10
When the Agilent 8702D finishes measuring the standard, connect the open
end of the open/short combination to reference test port 2.
11
Press
REVERSE:OPEN
.
12
When the Agilent 8702D finishes measuring the standard, connect the short
end of the open/short combination to reference test port 2.
13
Press
REVERSE:SHORT
.
14
When the Agilent 8702D finishes measuring the standard, connect the 50 ohm
termination to reference test port 2.
15
Press
REVERSE:LOAD
.
16
When the Agilent 8702D finishes measuring the standard, press
STANDARDS
DONE
. The Agilent 8702D briefly displays,
COMPUTING CAL COEFFICIENTS
.