2-4
Automated Verification
Step 2. Measurement Calibration
3
Insert the verification kit disk into the Agilent 8702Ds disk drive.
4
Press
PRESET
,
SAVE/RECALL
,
SELECT DISK
,
INTERNAL DISK
.
5
If you want a printout of the verification data for all the devices, press
SYSTEM
,
SERVICE MENU
,
TEST OPTIONS
,
RECORD ON
.
If you switch on the record function, you
cannot
switch it off during the verifi-
cation procedure.
6
Position the paper in the printer so that printing starts at the top of the page.
7
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
SYS VER TESTS
,
EXECUTE TEST
.
8
When the Agilent 8702D displays,
Sys Ver Init DONE
, the initialization
procedure is complete.
Do
not
press
PRESET
or recall another instrument state. You must use the in-
strument state that you loaded during the initialization procedure.
Step 2. Measurement Calibration
1
Press
CAL
,
CAL KIT & STDS
,
SELECT CAL KIT
,
CAL KIT:7mm
,
RETURN
,
RETURN
,
CALIBRATE
MENU
,
FULL 2-PORT
.
2
Press
ISOLATION
,
OMIT ISOLATION
,
ISOLATION DONE
.
3
Press
REFLECTION
.
4
Connect the open end of the open/short combination (supplied in the
calibration kit) to reference test port 1, as shown in Figure 2-2.