Agilent Technologies 87020 Stroller User Manual


 
4-41
Performance Tests
10. System Trace Noise (Only for Analyzers without Option 006)
2
Press
PRESET
,
MENU
,
POWER
,
5
,
x1
.
3
Press
RETURN
,
CW FREQ
,
3
,
G/n
,
NUMBER of POINTS
,
1601
,
x1
.
4
Press
MARKER FCTN
,
MARKER MODE MENU
,
STATS ON
to activate the instruments
statistic feature.
System Trace Noise for A/R Magnitude
5
Press
MEAS
,
Trans: E/E S12 REV
.
6
Press
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
7
When the analyzer displays the
Hld
annotation, press
SCALE REF
,
AUTO SCALE
.
8
Write the s.dev (standard deviation) value, which appears on the analyzer
display, on the Performance Test Record.
System Trace Noise for A/R Phase
9
Press
FORMAT
,
PHASE
.
10
Press
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
11
When the analyzer finishes the number of sweeps, press
SCALE REF
,
AUTO SCALE
.
12
Write the s.dev (standard deviation) value on the Performance Test Record.
System Trace Noise for B/R Magnitude
13
Press
MEAS
,
Trans: E/E S21 FWD
.
14
Press
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
15
When the analyzer finishes the number of sweeps, press
SCALE REF
,
AUTO SCALE
.
16
Write the s.dev (standard deviation) value on the Performance Test Record.
System Trace Noise for B/R Phase
17
Press
FORMAT
,
PHASE
.
18
Press
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
19
When the analyzer finishes the number of sweeps, press
SCALE REF
,
AUTO SCALE
.
20
Write the s.dev (standard deviation) value on the Performance Test Record.