4-26
Performance Tests
7. Test Port Input Frequency Response
Figure 4-14. Setup for Power Meter Calibration on Test Port 2
33
Press
START
,
3
,
G/n
,
STOP
,
6
,
G/n
.
34
Press
CAL
,
PWRMTR CAL
.
35
Press
LOSS/SENSR LISTS
,
CAL FACTOR SENSOR B
. Repeat Step 10 to build a
calibration factor sensor table for the Agilent 8481A power sensor.
36
Press
DONE
to exit the sensor calibration factor entries menu.
37
To select the Agilent 8481A power sensor, press
USE SENSOR B
.
38
Press
RETURN
,
TAKE CAL SWEEP
to start the power meter calibration.