Agilent Technologies 87020 Stroller User Manual


 
6-35
Performance Tests Option 011
11. Receiver Input Impedance
11. Receiver Input Impedance
Perform this test to verify the return loss of each of the analyzers receiver
inputs.
Required
Equipment
Test set, 50
. . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 85044A/85046A/85047A
2-way Power Splitter, 50
. . . . . . . . . . . . . . . . . . .Agilent 11667A Option 001
Attenuator, 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8491A Option 010
Termination, 50
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 908A
Calibration kit, 50
, Type-N. . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 85032B
Adapter, APC-7 to Type-N (m) . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11525A
RF cable set, 50
, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11851B
RF cable, 50
, 7 mm . . . . . . . . . . . . . . . . . . . . . . . . . . . Part of Agilent 11857D
Procedure
1
Connect the equipment as shown in Figure 6-17.
If you are using the Agilent 85044A test set, set the attenuation to 0 dB.
If you are using the Agilent 85046A or 85047A test set, connect the test set
interface cable to the analyzer and use the test set port 1.
Figure 6-17. Receiver B Input Impedance Test Setup