6-31
Performance Tests – Option 011
9. Receiver Input Crosstalk
A into B Crosstalk from 1 GHz to 3 GHz
38
Press
START
,
1
,
G/n
,
STOP
,
3
,
G/n
.
39
Press
MEAS
,
INPUT PORTS
,
A/R
,
AVG
,
AVERAGING OFF
,
MENU
,
TRIGGER MENU
,
SINGLE
.
40
At the end of the sweep, press
DISPLAY
,
DATA →MEMORY
,
DATA/MEM
,
MEAS
,
INPUT
PORTS
,
B/R
.
41
Press
AVG
,
AVERAGING FACTOR
,
5
,
x1
,
AVERAGING ON
,
MENU
,
TRIGGER MENU
,
NUMBER
of GROUPS
,
5
,
x1
.
42
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”
(Only for Option 006 Analyzers) A into B Crosstalk from 3 GHz to
4.5 GHz
43
Press
START
,
3
,
G/n
,
STOP
,
4.5
,
G/n
.
44
Press
MEAS
,
INPUT PORTS
,
A/R
,
AVG
,
AVERAGING OFF
,
MENU
,
TRIGGER MENU
,
SINGLE
.
45
At the end of the sweep, press
DISPLAY
,
DATA →MEMORY
,
DATA/MEM
,
MEAS
,
INPUT
PORTS
,
B/R
.
46
Press
AVG
,
AVERAGING FACTOR
,
5
,
x1
,
AVERAGING ON
,
MENU
,
TRIGGER MENU
,
NUMBER
of GROUPS
,
5
,
x1
.
47
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”
(Only for Option 006 Analyzers) A into B Crosstalk from 4.5 GHz to
6 GHz
48
Press
START
,
4.5
,
G/n
,
STOP
,
6
,
G/n
.
49
Press
MEAS
,
INPUT PORTS
,
A/R
,
AVG
,
AVERAGING OFF
,
MENU
,
TRIGGER MENU
,
SINGLE
.
50
At the end of the sweep, press
DISPLAY
,
DATA →MEMORY
,
DATA/MEM
,
MEAS
,
INPUT
PORTS
,
B/R
.
51
Press
AVG
,
AVERAGING FACTOR
,
5
,
x1
,
AVERAGING ON
,
MENU
,
TRIGGER MENU
,
NUMBER
of GROUPS
,
5
,
x1
.
52
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”