4-3
Performance Tests
1. Test Port Output Frequency Range and Accuracy
WARNING
Any servicing, adjustment, maintenance, or repair of this product
must be performed only by qualified personnel. Repair information
provided in the In Case of Difficulty sections of this chapter are
provided for qualified service personnel and are intended to be used
with the Agilent 8753D Service Guide.
1. Test Port Output Frequency Range and Accuracy
Perform this test to verify the frequency accuracy of the Agilent 8702D over
its entire operating frequency range.
Required
Equipment
Frequency Counter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 5343A
Cable, 50
Ω,
Type-N, 24-inch . . . . . . . . . . . . . . . . . . . . Agilent P/N 8120-4781
Adapter, APC-3.5 (f) to Type-N (f) . . . . . . . . . . . . . . . Agilent P/N 1250-1745
Adapter, APC-7 to Type-N (f) . . . . . . . . . . . . . . . . . . . . . .Agilent P/N 11524A
Adapter, Type-N (f) to BNC (m) . . . . . . . . . . . . . . . . . Agilent P/N 1250-1477
Additional Equipment for Agilent 8702D Option 075:
Minimum Loss Pad, 50
Ω
to 75
Ω
. . . . . . . . . . . . . . . . . . . . . . . Agilent 11852B
Procedure
1
Connect the equipment as shown in Figure 4-1.
Agilent 8753D is shown in the figures
The procedures in this chapter are designed to work with Agilent 8702D and
Agilent 8753D instruments. Because of this, many of the figures label the instrument
being tested as an Agilent 8753D. Simply substitute your Agilent 8702D for the
Agilent 8753D.