Performance Tests Option 011
4. Receiver Minimum R Channel Level for External Source Mode
5
Refer to the
Agilent 8753D Option 011 Service Guide
for more
troubleshooting information.
4. Receiver Minimum R Channel Level for External
Source Mode
Perform this test to verify proper phase lock in the external source mode, as
the specified minimum R input level of 25 dBm.
Required
Equipment
External source. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 83640A
Adapter, APC-3.5 (f) to Type-N (f). . . . . . . . . . . . . . . .Agilent P/N 1820-1745
Cable, 50
, Type-N . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11851B
Attenuator, 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 010
Attenuator, 20 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 020
Procedure
1
Connect the equipment as shown in Figure 6-6.
Figure 6-6. Receiver Minimum R Channel Level for External Source Test Setup