Performance Tests Option 011
12. Receiver Magnitude Dynamic Accuracy
6
On the Agilent 8902A, press the blue shift key and
SET REF
.
7
On the Agilent 8702D Option 011:
a
Press
CAL
,
CALIBRATE MENU
,
RESPONSE
,
THRU
.
b
After the analyzer finishes the measurement, press
DONE: RESPONSE
.
c
Press
SAVE/RECALL
,
SAVE STATE
.
d
Press
MARKER FCTN
,
MARKER MODE MENU
,
STATS ON
.
8
Set the Agilent 8496A to 0 dB.
9
On the Agilent 8702D Option 011, press
MENU
,
TRIGGER MENU
,
SINGLE
.
10
Write the mean value (which appears on the analyzers display) in the
"Receiver Channel A" column of the "Performance Test Record."
11
Write the Agilent 8902A readout in the "8902A" column of the "Performance
Test Record."
12
Calculate the difference between the "Receiver Channel A" and the "8902A".
13
Take the absolute value of the calculated difference, and enter the result in the
"Magnitude Value" column of the "Performance Test Record."
14
Set the Agilent 8496A to the other attenuation settings listed in the
"Performance Test Record."
15
On the Agilent 8702D Option 011, press
SINGLE
. Repeat Step 9 through Step 13.
Channel B Magnitude Dynamic Accuracy
16
Set the Agilent 8496A to 20 dB.
17
On the Agilent 8702D Option 011, press
CONTINUOUS
,
MEAS
,
INPUT PORTS
,
B
.
NOTE
Ignore the
UNCAL
displayed on the Agilent 8902A. This indicates that you are making
an approximate power level measurement. However, this is
not
a concern because you
will be making a relative power measurement.
NOTE
When the Agilent 8902A displays the "
RECAL
" annunciation, press CALIBRATE.