Installing the Agilent 8702D
Step 14. Copy the EEPROM Disk
Step 14. Copy the EEPROM Disk
The Agilent 8702D is shipped from the factory with an EEPROM backup disk
which is unique to each instrument. The backup disk contains a copy of cor-
rection constants which were stored in the instruments internal memory at
the factory. If your Agilent 8702D should ever be damaged, these correction
constants may need to be reinstalled from the disk. Use the following proce-
dure to create a backup copy of the disk.
1
Insert a 3.5-inch disk into the Agilent 8702Ds disk drive.
2
Press
PRESET
so that the preset state is stored.
3
If the disk is not formatted, press
SAVE/RECALL
,
FILE
UTILITIES
,
FORMAT
DISK
.
To format a LIF disk, select
FORMAT:LIF
(The supplied EEPROM backup disk
is LIF. The Agilent 8702D does not support LIF-HFS format.)
To format a DOS disk, select
FORMAT:DOS
.
Press
FORMAT INT DISK,
and answer
YES
at the query.
4
Press
SYSTEM
,
SERVICE
MENU
,
SERVICE
MODES
,
MORE
,
STORE
EEPR
ON
,
SAVE/RECALL,
SAVE RCL MENU
,
SELECT
DISK
,
INTERNAL
DISK
,
RETURN
,
SAVE
STATE
to store the
correction-constants data onto floppy disk.
A default file FILE0 is created. The file name appears in the upper left-hand
corner of the display. The file type ISTATE(E) describes the file as an instru-
ment-state with EEPROM backup.
5
Press
FILE
UTILITIES
,
RENAME
FILE
,
ERASE
TITLE
. Use the RPG and the
SELECT
LETTER
softkey to rename the file FILE0 to N12345 where 12345 represents the last
5 digits of the instruments serial number. (The first character in the file name
must be a letter.) When finished, press
DONE
.
6
Label the disk with the serial number of the instrument and the words
EEPROM Backup Disk.
Whenever the Agilent 8702D is returned to Agilent Technologies for servicing
and/or calibration, the EEPROM backup disk should be returned with the
Agilent 8702D. This will significantly reduce the instrument repair time. The
EEPROM backup disk procedure is now complete.