5-20
Automated Verification – Option 011
In Case of Difficulty
8
When the analyzer finishes all the measurements, reconnect the mismatch
verification device, as shown in Figure 5-14.
Figure 5-14. Mismatch Device Verification Setup
9
Press
RETURN
,
TESTS
,
31
,
x1
,
EXECUTE TEST, CONTINUE
.
10
You have completed the system verification procedure when the analyzer
displays,
Ver Def 4 PASS
.
In Case of Difficulty
1
Inspect all connections. Do
not
disconnect the cables from the analyzer test
ports. Doing so will invalidate the calibration that you have done earlier.
2
Recall the measurement calibration and repeat the procedure:
a
Press
PRESET
,
SAVE/RECALL
,
SELECT DISK
,
INTERNAL DISK
,
RETURN
.
b
Using the front panel knob, highlight the title of the one-path 2-port calibra-
tion that you have done earlier.