Agilent Technologies 87020 Stroller User Manual


 
4-83
Performance Tests
Performance Test Records
Test 13. Test Port Receiver Magnitude Dynamic Accuracy
Test Port
Input Power
8496A
Setting
(dB)
Test Port
(dB)
8902A
(dB)
Calculated
Value (dB)
Spec. (dB)
Mea.
Uncer. (dB)
Test Port 2
10 0
_________ _________ _________
0.059
± 0.01
20 10
_________ _________ _________
0.050
± 0.02
30 20
_________ _________ _________
0.050
± 0.02
40 30
_________ _________ _________
0.053
± 0.04
50 40
_________ _________ _________
0.058
± 0.06
60 50
_________ _________ _________
0.095
± 0.08
70 60
_________ _________ _________
0.250
± 0.14
80 70
_________ _________ _________
0.750
± 0.16
90 80
_________ _________ _________
2.200
± 0.18
100 90
_________ _________ _________
5.200
± 0.20
Test Port 1
10 0
_________ _________ _________
0.059
± 0.01
20 10
_________ _________ _________
0.050
± 0.02
30 20
_________ _________ _________
0.050
± 0.02
40 30
_________ _________ _________
0.053
± 0.04
50 40
_________ _________ _________
0.058
± 0.06
60 50
_________ _________ _________
0.095
± 0.08
70 60
_________ _________ _________
0.250
± 0.14
80 70
_________ _________ _________
0.750
± 0.16
90 80
_________ _________ _________
2.200
± 0.18
100 90
_________ _________ _________
5.200
± 0.20