6-29
Performance Tests – Option 011
9. Receiver Input Crosstalk
B into A Crosstalk from 300 kHz to 1 GHz
22
Connect the equipment as shown in Figure 6-14.
Figure 6-14. B into A Receiver Input Crosstalk Test Setup
23
Press
START
,
300
,
k/m
,
STOP
,
1
,
G/n.
24
Press
MEAS
,
INPUT PORTS
,
A/B
,
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
25
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”
B into A Crosstalk from 1 GHz to 3 GHz
26
Press
START
,
1
,
G/n
,
STOP
,
3
,
G/n
,
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
27
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the “Performance Test Record.”
(Only for Option 006 Analyzers) B into A Crosstalk from 3 GHz to
4.5 GHz
28
Press
START
,
3
,
G/n
,
STOP
,
4.5
,
G/n
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
29
When the analyzer finishes the number of sweeps, write the marker value,