Agilent Technologies 87020 Stroller User Manual


 
6-59
Performance Tests Option 011
15. Receiver Phase Compression
keypad to enter the Start Power value as determined in Step 18.
21
Press
STOP
and enter the Stop Power value as determined in Step 19.
22
Press
MENU
,
TRIGGER MENU
,
SINGLE
.
23
At the end of the sweep, press
SCALE REF
,
AUTO SCALE
.
24
Press
MARKER
,
MODE MENU
,
REF=1
,
MARKER FCTN
,
MARKER MODE MENU
,
MKR
SEARCH
,
SEARCH: MAX
.
25
Press
MARKER
,
MARKER 2
,
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH: MIN
.
26
Write the absolute value of marker 2 in the Measured Value column on the
"Performance Test Record."
27
Repeat Step 16 through Step 25 for the other frequencies listed on the
Performance Test Record.