6-52
Performance Tests Option 011
14. Receiver Magnitude Compression
20
Press
STOP
and enter the Stop Power value as determined in Step 18.
21
Press
MENU
,
TRIGGER MENU
,
SINGLE
.
22
At the end of the sweep, press
SCALE REF
,
AUTO SCALE
.
23
Press
MARKER
,
MODE MENU
,
REF=1
,
MARKER FCTN
,
MARKER MODE MENU
,
MKR
SEARCH
,
SEARCH: MAX
.
24
Press
MARKER
,
MARKER 2
,
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH: MIN
.
25
Write the absolute value of marker 2 in the Measured Value column on the
"Performance Test Record."
26
Repeat Step 15 through Step 25 for the other frequencies listed on the
Performance Test Record.