Agilent Technologies 87020 Stroller User Manual


 
6-5
Performance Tests Option 011
2. Source Power Range, Linearity, and Accuracy
2. Source Power Range, Linearity, and Accuracy
Perform this test to verify the analyzers power range and linearity at different
CW frequencies throughout the analyzer operating frequency range.
Required
Equipment
Power Meter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 436A/437B/438A
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8482A
2-way Power Splitter (2), 50
. . . . . . . . . . . . . . . .Agilent 11667A Option 001
Attenuator (2), 20 dB . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8491A Option 020
RF Cable Set, 50
, Type-N. . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 11851B
Additional Required Equipment for Analyzers with Option 006
Power Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8481A
Procedure Path loss calibration
1
Connect the equipment as shown in Figure 6-2.
Figure 6-2. Path Loss Calibration Test Setup (#1)