6-81
Performance Tests – Option 011
Performance Test Records - Option 011
Test 12. Receiver Magnitude Dynamic Accuracy (1 of 2)
Receiver
Input Power (dB)
Agilent
8496A
Setting
(dB)
Receiver
Channel A
(dB)
Agilent
8902A
(dB)
Magnitude
Value (dB)
Spec. (dB)
Meas.
Uncer. (dB)
Channel A
– 10 0
_________ _________ _________
≤
0.075
± 0.010
– 20 10
_________ _________ _________
≤
0.059
± 0.020
– 30 20
_________ _________ _________
≤
0.050
± 0.020
– 40 30
_________ _________ _________
≤
0.050
± 0.040
– 50 40
_________ _________ _________
≤
0.050
± 0.060
– 60 50
_________ _________ _________
≤
0.058
± 0.080
– 70 60
_________ _________ _________
≤
0.089
± 0.140
– 80 70
_________ _________ _________
≤
0.240
± 0.160
– 90 80
_________ _________ _________
≤
0.680
± 0.180
– 100 90
_________ _________ _________
≤
1.950
± 0.200
– 110 100
_________ _________ _________
≤
5.200
± 0.280
Channel B
– 10 0
_________ _________ _________
≤
0.075
± 0.010
– 20 10
_________ _________ _________
≤
0.059
± 0.020
– 30 20
_________ _________ _________
≤
0.050
± 0.020
– 40 30
_________ _________ _________
≤
0.050
± 0.040
– 50 40
_________ _________ _________
≤
0.050
± 0.060
– 60 50
_________ _________ _________
≤
0.058
± 0.080
– 70 60
_________ _________ _________
≤
0.089
± 0.140
– 80 70
_________ _________ _________
≤
0.240
± 0.160