4-54
Performance Tests
13. Test Port Receiver Magnitude Dynamic Accuracy
19
On the Agilent 8902A, press the blue shift key and
SET REF
.
20
On the Agilent 8702D:
a
Press
CAL
,
CALIBRATE MENU
,
RESPONSE
,
THRU
.
b
After the analyzer finishes the measurement, press
DONE: RESPONSE
.
c
Press
SAVE/RECALL
,
SAVE STATE
.
21
Set the Agilent 8496A to 0 dB.
22
On the Agilent 8702D, press
MENU
,
TRIGGER MENU
,
SINGLE
.
23
Write the mean value in the “Test Port” column of the “Performance Test
Record.”
24
Write the Agilent 8902A readout in the “8902A” column of the “Performance
Test Record.”
25
Calculate the difference between the “Test Port 1" and the “8902A”.
26
Take the absolute value of the calculated difference and enter the result in the
“Calculated Value” column of the “Performance Test Record.”
27
Set the Agilent 8496A to the other attenuation settings listed in the
“Performance Test Record,” while repeating Step 22 through Step 26.
NOTE
When the Agilent 8902A displays the “
RECAL
” annunciation, press CALIBRATE.