Agilent Technologies 87020 Stroller User Manual


 
6-13
Performance Tests Option 011
4. Receiver Minimum R Channel Level for External Source Mode
2
On the external source, press
PRESET
,
POWER LEVEL
,
4
,
dB(m)
,
CW
,
10
,
MHz
.
3
On the analyzer, press
PRESET
,
MENU
,
CW FREQ
,
10
,
M/
µ
,
SYSTEM
,
INSTRUMENT
MODE
,
EXT SOURCE AUTO
to set up the analyzer for an external source input to
the receiver channel R.
4
Press
SYSTEM
,
SERVICE MENU
,
ANALOG BUS ON
,
MEAS
,
S PARAMETERS
,
ANALOG IN
Aux Input
,
29
,
x1
,
COUNTER: FRAC N
,
MEAS
,
INPUT PORTS
,
R
, to set up the analog bus
counter to measure the frequency of the fractional-N VCO.
5
Press
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
TRACKING ON
,
SEARCH:MAX
to observe the maximum value of the receiver R input signal.
6
The analyzer marker 1 should read 26 dBm (which appears on the analyzer
display). If the marker value exceeds 26 dBm, press
POWER LEVEL
on the
external source and decrease the power until the marker 1 reads 26 dBm.
7
Press
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH:OFF
.
8
Write the analog bus counter reading (which appears on the analyzer display)
on the Performance Test Record.
9
On the external source, press
CW
,
20
,
MHz
.
10
On the analyzer, press
MENU
,
CW FREQ
,
20
,
M
/
µ
. If the analyzers marker value
readout exceeds 26 dBm, reduce the external source power level. Write the
analog bus counter reading and compare it with the acceptable limits on the
Performance Test Record.
11
Repeat Step 9 and Step 10 for the CW frequencies listed on the Performance
Test Record.
NOTE
The analyzer is phase-locked if the analog bus counter reading is within the limits shown
on the Performance Test Record.