Agilent Technologies 87020 Stroller User Manual


 
4-70
Performance Tests
18. Test Port Output/Input Harmonics (Option 002 Analyzers with Option 006 Only)
harmonic.
46
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
.
47
After one sweep, press
DISPLAY
,
DATAMEMORY
,
DATA/MEM
to normalize the
trace.
48
Press
SCALE REF
,
AUTO SCALE
,
SCALE/DIV
,
1
,
x1
to get a better viewing of the trace.
49
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC THIRD
.
50
After one sweep, press
SCALE REF
,
AUTO SCALE
.
51
Press
MARKER FCTN
,
MARKER MODE MENU
,
MKR SEARCH
,
SEARCH: MAX
.
52
Write the marker 1 value on the Performance Test Record.