6-45
Performance Tests – Option 011
12. Receiver Magnitude Dynamic Accuracy
Figure 6-24. Channel R Magnitude Dynamic Accuracy Test Setup
31
Set the Agilent 8496A to 20 dB.
32
On the Agilent 8702D Option 011:
a
Press
CONTINUOUS
,
MEAS
,
INPUT PORTS
,
R
,
MENU
,
POWER
.
b
Using the analyzer front panel knob, adjust the test port power for a reading
of 0.000 dB on the Agilent 8902A.
c
Wait for the reading to stabilize.
33
On the Agilent 8902A, press the blue shift key and
SET REF
.
34
On the Agilent 8702D Option 011:
a
Press
CAL
,
CALIBRATE MENU
,
RESPONSE
,
THRU
.
b
After the analyzer finishes the measurement, press
DONE: RESPONSE
.
c
Press
SAVE/RECALL
,
SAVE STATE
.
35
Set the Agilent 8496A to 0 dB.