4-22
Performance Tests
7. Test Port Input Frequency Response
a
From the sensor calibration factor entries menu, press
ADD
.
b
Press
FREQUENCY
,
100
,
k/m
. If you make an entry error, press
and re-enter
the correct value again.
c
Press
CAL FACTOR
,
96.4
,
x1
.
d
Press
DONE
to terminate the first calibration factor data point entry.
e
To enter the second cal factor data point, press
ADD
.
f
Press
FREQUENCY
,
300
,
k/m
.
g
Press
CAL FACTOR
,
98.4
,
x1
.
h
To terminate the second calibration factor data point entry, press
DONE
.
i
Press
SEGMENT
and use the front panel knob to scroll through the sensor cal-
ibration factors table. Check to be sure all values are entered correctly. If
you spot an error, use the front panel knob to point to the data point you
want to modify and press
EDIT
.
13
Press the appropriate softkeys to create a power sensor calibration factors
table.
14
Press
DONE
to exit the sensor calibration factor entries menu.
15
Press
RETURN
,
ONE SWEEP, TAKE CAL SWEEP
to start the power meter calibration.
Wait until the analyzer finishes the sweep, then continue with this procedure.
NOTE
The analyzer displays the
PC
annotation, indicating the power meter calibration is done
and the error correction is active.