Agilent Technologies 87020 Stroller User Manual


 
3-8
Manual Verification
Step 3. Device Verification
Figure 3-5. Mismatch Device Verification Setup
17
Press
TESTS
,
30
,
x1
,
EXECUTE TEST, CONTINUE
.
Ignore the error message.
18
Press
MENU
,
POWER,
2
,
x1
,
CAL
,
CORRECTION ON
.
19
Repeat Step 5 through Step 8 until you have made all the S-parameter
measurements for the verification device.
20
When the Agilent 8702D displays,
Ver Dev 3 FAIL
, proceed to the next step
only if
all
S-parameter measurements show a
PASS
status under the same
condition mentioned in Step 10.
21
Connect the verification mismatch as shown in Figure 3-6. Notice that port B
of the verification mismatch is now connected to the reference test port 1.