Agilent Technologies 87020 Stroller User Manual


 
6-28
Performance Tests Option 011
9. Receiver Input Crosstalk
(Only for Option 006 Analyzers) R into A Crosstalk from 3 GHz to
4.5 GHz
9
Press
START
,
3
,
G/n
,
STOP
,
4.5
,
G/n
,
NUMBER of GROUPS
,
5
,
x1
.
10
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.
(Only for Option 006 Analyzers) R into A Crosstalk from 4.5 GHz to
6 GHz
11
Press
START
,
4.5
,
G/n
,
STOP
,
6
,
G/n
,
NUMBER of GROUPS
,
5
,
x1
.
12
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.
R into B Crosstalk from 300 kHz to 1 GHz
13
Press
START
,
300
,
k/m
,
STOP
,
1
,
G/n.
14
Press
MEAS
,
INPUT PORTS
,
B/R
,
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
15
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.
R into B Crosstalk from 1 GHz to 3 GHz
16
Press
START
,
1
,
G/n
,
STOP
,
3
,
G/n
,
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
17
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.
(Only for Option 006 Analyzers) R into B Crosstalk from 3 GHz to
4.5 GHz
18
Press
START
,
3
,
G/n
,
STOP
,
4.5
,
G/n
,
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
19
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.
(Only for Option 006 Analyzers) R into B Crosstalk from 4.5 GHz to
6 GHz
20
Press
START
,
4.5
,
G/n
,
STOP
,
6
,
G/n
,
MENU
,
TRIGGER MENU
,
NUMBER of GROUPS
,
5
,
x1
.
21
When the analyzer finishes the number of sweeps, write the marker value,
which appears on the analyzer display, on the Performance Test Record.