Performance Tests Option 011
16. Source and Receiver Harmonics (Option 002 Only)
24
Move the receiver A input cable to receiver input B.
25
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
,
STOP
,
1.5
,
G/n
.
26 Only for Analyzers with Option 006:
Press
STOP
,
3
,
G/n
.
27
Repeat Step 6 through Step 9 to measure the source and receiver worst case
2nd harmonic for the receiver input B. Press
MEAS
,
INPUT PORTS
,
B
in Step 6.
28
Repeat Step 11 through Step 16 to measure the source and receiver worst case
3nd harmonic for the receiver input B.
Receiver Harmonics
29
Connect the equipment as shown in Figure 6-34.
Figure 6-34. Receiver Harmonics Test Setup
30
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
,
STOP
,
1.5
,
G/n
.
31 Only for Analyzers with Option 006:
Press
STOP
,
3
,
G/n
.
32
Repeat Step 6 through Step 9 to measure the receiver worst case 2nd harmonic
for the receiver input B. Press
MEAS
,
INPUT PORTS
,
B
in Step 6.
33
Repeat Step 11 through Step 16 to measure the receiver worst case 3nd
harmonic for the receiver input B.
34
Press
SYSTEM
,
HARMONIC MEAS
,
HARMONIC OFF
.
35
Repeat Step 6 through Step 9 to measure the receiver worst case 2nd harmonic