Agilent Technologies 87020 Stroller User Manual


 
5-11
Automated Verification Option 011
In Case of Difficulty
8
When the analyzer finishes all the measurements, reconnect the mismatch
verification device, as shown in Figure 5-7.
Figure 5-7. Mismatch Device Verification Setup
9
Press
RETURN, TESTS
,
31
,
x1
,
EXECUTE TEST, CONTINUE
.
10
You have completed the system verification procedure when the analyzer
displays,
TEST 31 Ver Dev 4 PASS
.
In Case of Difficulty
1
Inspect all connections. Do
not
disconnect the cables from the analyzer test
ports. Doing so will invalidate the calibration that you have done earlier.
2
Press
PRESET
,
SAVE/RECALL
.
a
Using the front panel knob, highlight the title of the full 2-port measurement
calibration that you have done earlier.
b
Press
RECALL STATE
.
c
Repeat Step 3. Device Verification on page 5-8.