1-7
Installing the Agilent 8702D
Step 4. Check the Fuse and Voltage Selection
Step 4. Check the Fuse and Voltage Selection
1
Locate the line-input connector on the instruments rear panel.
2
Disconnect the line-power cable if it is connected.
3
Use a small flat-blade screwdriver to open the pull-out fuse drawer.
WARNING
The power cord is connected to internal capacitors that may remain
live for 10 seconds after you disconnect the cord from the power
supply.
Figure 1-3. Line Fuse Removal and Replacement
4
Verify that the value of the line-voltage fuse in the pull-out drawer is correct.
The recommended fuse is an IEC 127 5
×
20 mm, 3A, 250 V, Agilent part number
2110-0780.
Notice that an extra fuse is provided in a drawer located on the fuse holder.