3-5
Manual Verification
Step 3. Device Verification
18
Connect the test port cables together as shown in Figure 3-2.
Figure 3-2. Transmission Calibration Setup
19
Press
TRANSMISSION
,
FWD TRANS THRU
.
20
When the Agilent 8702D finishes measuring the standard, press
FWD MATCH
THRU
.
21
When the Agilent 8702D finishes measuring the standard, press
REV TRANS
THRU
.
22
When the Agilent 8702D finishes measuring the standard, press
REV MATCH
THRU
.
23
When the Agilent 8702D finishes measuring the standard, press
STANDARDS
DONE
,
DONE 2-PORT CAL
.
24
Press
SAVE/RECALL
,
SELECT DISK
,
INTERNAL MEMORY
,
RETURN
,
SAVE STATE
to save
the measurement calibration into the Agilent 8702D internal memory.
25
Press
SELECT DISK
,
INTERNAL DISK
.
Step 3. Device Verification
1
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
28
,
x1
,
EXECUTE TEST
.
2
Connect the 20 dB attenuator (supplied in the verification kit) as shown in