4-21
Performance Tests
7. Test Port Input Frequency Response
8
Press
MENU
,
NUMBER of POINTS
,
51
,
x1
.
9
Press
POWER
,
PWR RANGE MAN
to turn the auto power range off.
10
Press
PORT POWER
to uncouple the test port output power.
11
Press
CAL
,
PWRMTR CAL
.
12
Press
LOSS/SENSR LISTS
,
CAL FACTOR SENSOR A
. Refer to the back of the power
sensor to locate the different calibration factor values along with their
corresponding frequencies.
The following softkeys are included in the sensor calibration factor entries
menu:
SEGMENT
press to select a point where you can use the front panel
knob or entry keys to enter a value
EDIT
press to edit or change a previously entered value
DELETE
press to delete a point from the sensor calibration factor
table
ADD
select this key to add a point into the sensor calibration
factor table
CLEAR LIST
select this key to erase the entire sensor calibration factor
table
DONE
select this key when done entering points to the sensor
calibration factor table
As an example, the following are the keystrokes for entering the first two cali-
bration factor data points for the Agilent 8482A power sensor (assuming
CF%= 96.4 at 100 kHz and CF%= 98.4 at 300 kHz):
NOTE
The analyzer displays the
PRm
annotation, indicating that the analyzer power range is
set to MANUAL.
NOTE
The analyzers calibration factor sensor table can hold a MAXIMUM of 12 calibration
factor data points.