Agilent Technologies 87020 Stroller User Manual


 
4-37
Performance Tests
9. Calibration Coefficients
Transmission Tracking (Forward) Calibration Coefficient
31
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
37
,
x1
,
EXECUTE TEST
.
32
When the analyzer finishes the test, repeat Step 25 through Step 28.
Reflection Tracking (Forward) Calibration Coefficient
33
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
34
,
x1
,
EXECUTE TEST
.
34
When the analyzer finishes the test, repeat Step 25 through Step 28.
Load Match (Reverse) Calibration Coefficient
35
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
42
,
x1
,
EXECUTE TEST
.
36
When the analyzer finishes the test, repeat Step 25 through Step 28.
Transmission Tracking (Reverse) Calibration Coefficient
37
Press
SYSTEM
,
SERVICE MENU
,
TESTS
,
43
,
x1
,
EXECUTE TEST
.
38
When the analyzer finishes the test, repeat Step 25 through Step 28.
Second Full 2-Port Calibration
39
Connect the equipment as shown in Figure 4-22.
Figure 4-22. Second Full 2-Port Calibration Test Setup