6-37
Performance Tests – Option 011
11. Receiver Input Impedance
Figure 6-18. Receiver A Input Impedance Test Setup
14
Press
PRESET
,
MEAS
,
INPUT PORTS
,
B/R
,
MENU
,
NUMBER of POINTS
,
1601
,
x1
.
15
Press
CAL
,
CALIBRATE MENU
,
S11 1-PORT
. Repeat Step 4 through Step 12 for the
analyzer receiver input A return loss.