Performance Tests
Performance Test Records
Test 14. Test Port Receiver Phase Dynamic Accuracy
Test Port
Input Power
Calculated
Value (dB)
A B
Results
Value
(degrees)
Spec.
(degrees)
Mea.
Uncer.
Test Port 2
10
_________ _________ _________ _________
0.42
N/A
20
_________ _________ _________ _________
0.35
N/A
30
_________ _________ _________ _________
0.35
N/A
40
_________ _________ _________ _________
0.36
N/A
50
_________ _________ _________ _________
0.39
N/A
60
_________ _________ _________ _________
0.64
N/A
70
_________ _________ _________ _________
1.75
N/A
80
_________ _________ _________ _________
5.00
N/A
90
_________ _________ _________ _________
17.50
N/A
100
_________ _________ _________ _________
53.00
N/A
Test Port 1
10
_________ _________ _________ _________
0.42
N/A
20
_________ _________ _________ _________
0.35
N/A
30
_________ _________ _________ _________
0.35
N/A
40
_________ _________ _________ _________
0.36
N/A
50
_________ _________ _________ _________
0.39
N/A
60
_________ _________ _________ _________
0.64
N/A
70
_________ _________ _________ _________
1.75
N/A
80
_________ _________ _________ _________
5.00
N/A
90
_________ _________ _________ _________
17.50
N/A
100
_________ _________ _________ _________
53.00