Agilent Technologies 87020 Stroller User Manual


 
5-9
Automated Verification Option 011
Step 3. Device Verification
3
Press
CONTINUE
to run the test:
If you switched the record function
off
, you have to press
CONTINUE
after
each S-parameter measurement.
If you switched the record function
on
, the test displays and prints the pass/
fail information for the S-parameter measurements that are valid for system
verification.
4
When the analyzer finishes all the measurements, connect the 50 dB attenuator
(supplied in the verification kit), as shown in Figure 5-5.
Figure 5-5. Connections for the 50 dB Verification Device
5
Press
RETURN
,
TESTS
,
29
,
x1
,
EXECUTE TEST
,
CONTINUE
.