Performance Tests Option 011
12. Receiver Magnitude Dynamic Accuracy
Figure 6-21. Receiver Magnitude Dynamic Accuracy
Required
Equipment
Measuring Receiver . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8902A
Step Attenuator, 110 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Agilent 8496A
Attenuator, 10 dB . . . . . . . . . . . . . . . . . . . . . . . . . . . Agilent 8491A Option 010
Cable (4), 50
, Type-N, 24-inch . . . . . . . . . . . . . . . . . .Agilent P/N 8120-4781
2-Way Power Splitter (2), 50
. . . . . . . . . . . . . . . . . . . . . . . . .Agilent 11667A
Adapter (2), Type-N (m) to Type-N (m) . . . . . . . . . . .Agilent P/N 1250-1475
Procedure Channel A Magnitude Dynamic Accuracy
1
On the Agilent 8902A:
a
Press the blue shift key and
INSTR PRESET
,
CLEAR
.
b
Press the yellow shift key and
TUNED RF LEVEL
.
c
Press
LOG/LIN
to have the Agilent 8902A displayed linear readings (dBm).
d
Press
32.1
,
SPCL
.
2
Set the Agilent 8496A to 20 dB.
3
On the Agilent 8702D Option 011:
a
Press
PRESET
,
MENU
,
CW FREQ
,
30
,
M/
µ
.
b
Press
NUMBER of POINTS
,
51
,
x1
.